• Benefits and Limitations of Plunger Protection Layer for Bottle Top Dispensers

Laboratory Products

Benefits and Limitations of Plunger Protection Layer for Bottle Top Dispensers

Jan 23 2015

Bottle top dispensers are widely used in laboratories for the repeat dispensing of reagents and chemicals, strait from the bottle. During use, the liquid comes in direct contact with plunger and barrel. Most suppliers offer models with and without plunger protection layer. Below review summarises benefits and limitations of a protection layer, thus providing users better awareness when choosing a new dispenser.

Common materials used are fluoropolymers (FPs), for example PTFE, FEP and PFA. Main physical-chemical properties of these materials are: very smooth surface, resistant to build-up biofilms, seamless finish giving no adhesion point for crystal building; oleophobic, hydrophobic, low binding capacities with liquids like water and water based solutions; low permeability to liquid; reduction of the coefficient of friction compare to uncoated glass surface; excellent chemical resistance; food contact compliance (FDA compatibility).

A protection layer prevents plunger freezing when liquid remains in the instrument. It is recommended when handling salt solutions, acids, bases and buffers.

A known limitation of protection layer is its mechanical fragility. Scratches may impair proper working of the dispenser. Sufficient coating thickness is a way to overcome this.  Organic components such as solvents and some strong acids in high concentration may react with the coating. This may lead to layer swelling or peeling in extreme cases.

Plungers are usually made of ground glass or ceramic. Both materials being inert from a chemical point of view, they are perfectly suitable to handle liquids that would potentially react with FP’s protection layer.

One difficulty when using ground glass is to guarantee a smooth glide of plunger in the barrel. Manufacturer must find the best compromise between appropriate surface roughness and capillarity effects. This is only possible when plungers are machined with very high precision. Other known limitations are: high friction observed with some liquids, higher sensitivity to freezing when dispenser not used, scratch of barrel wall if forced is applied on plunger or if particles are stocked between plunger and barrel, increased capillary effects and ‘sticking’ feeling during plunger activation.

Ceramic (such as Al2O3) shows excellent chemical resistance and a better mechanical shock resistance compared to glass. But it is a highly expensive material increasing the market price of the instrument.

In conclusion, it is important for the Lab users to know what they intend to distribute before selection a suitable dispenser. Manufacturer must be clear about recommended use and limitations of their instruments. In any case, manufacturers must remain anytime available to provide all necessary technical support when needed.

See complete Socorex liquid handling programme at: www.socorex.com


Digital Edition

Lab Asia 31.2 April 2024

April 2024

In This Edition Chromatography Articles - Approaches to troubleshooting an SPE method for the analysis of oligonucleotides (pt i) - High-precision liquid flow processes demand full fluidic c...

View all digital editions

Events

FORUMESURE

Apr 22 2024 Marrakech, Morroco

Korea Lab 2024

Apr 23 2024 Kintex, South Korea

Korea Chem 2024

Apr 23 2024 Seoul, South Korea

Lab Indonesia

Apr 24 2024 Jakarta, Indonesia

View all events