Analytical Instrumentation

A Safe and Fast Solution for Accurate Quantification of Heavy Residues in LPG by Gas Chromatography: Representative Liquefied Gas Sample Introduction via High Pressure On-Column Injection into a Gas Chromatographic System - Lenny Kouwenhoven & Ani

Sep 15 2010

Author: Lenny Kouwenhoven & Anita Ruissen on behalf of Unassigned Independent Article

Free to read

This article has been unlocked and is ready to read.

Download

One of the specifications of commercial LPG is the amount of oily residue. Oily residue leads to troublesome deposits that will accumulate and corrode or plug the LPG fuel filter, the pressure regulators, the fuel mixer or
the control solenoids. Currently ASTM D2158 or ISO 13757 are used. These methods main disadvantage is that large amounts of LPG need to be evaporated which is a safety risk. The preferred alternative technique is gas
chromatography. However, the quantitative analysis of high boiling components in low boiling matrices is an analytical challenge. The limitation is the sample introduction. Residue adsorption on transfer lines, vaporisers
and valves lead to cross sample contamination or quantitative bias due to the fractionation of the heavy material. A new liquefied gas injector is developed by Da Vinci Europe to inject under pressure directly onto the
column. The sample remains in liquid phase, at room temperature and without contact with transfer lines, vaporisers or valves and therefore overcomes all limitations of the conventional sample introduction techniques.
The chromatography after this representative sample introduction is based on boiling point separation and the total amount is reported in parts per million.

Control over the residue content is essential in end-use applications of automotive LPG. These residues can lead to troublesome deposits that will accumulate and corrode or plug the LPG fuel filter, the low pressure
regulators, the fuel mixer or the control solenoids.

LPG can become contaminated by oily residues during its production or transport. Transport contamination can come from shared pipelines, valves and trucks used for the distribution of other products. Production sources such as the desulfurisation process may contribute sulfur absorbent oil to the LPG stream.

Commercial LPG, especially automotive applications, should comply with current fuel specifications for oily residue as described in ASTM Method D2158[1] or ISO 13757[2]. ASTM Method D2158 is commonly called oil stain method. After evaporating 100 milliliters of the LPG stream, the residue is dissolved in a solvent and the resulting solution is slowly dripped on the adsorption paper. The size and persistence of the stain which
remains on the paper after the solvent evaporates is an empirical measure of the oily residue in the LPG sample.

The ISO 13757 method quantifies the oily residue by weighing what remains after evaporation of two kilograms of the LPG. Both of these procedures offer an undue safety risk due to the required evaporation step
into the immediate environment.

Free to read

This article has been unlocked and is ready to read.

Download


Digital Edition

Lab Asia 31.2 April 2024

April 2024

In This Edition Chromatography Articles - Approaches to troubleshooting an SPE method for the analysis of oligonucleotides (pt i) - High-precision liquid flow processes demand full fluidic c...

View all digital editions

Events

Expomed Eurasia

Apr 25 2024 Istanbul, Turkey

AOCS Annual Meeting & Expo

Apr 28 2024 Montreal, Quebec, Canada

SETAC Europe

May 05 2024 Seville, Spain

InformEx Zone at CPhl North America

May 07 2024 Pennsylvania, PA, USA

ISHM 2024

May 14 2024 Oklahoma City, OK, USA

View all events