• Analysis of oxidised monoclonal antibodies using YMC's BioPro HIC BF

Bioanalytical

Analysis of oxidised monoclonal antibodies using YMC's BioPro HIC BF

Sep 22 2020

During the manufacture and/or storage of biopharmaceuticals, variants with different properties from desired substances are produced by enzyme reactions or physicochemical interactions. Characterisation of these variants is of great importance from the perspective of ensuring efficacy and safety of pharmaceutical products.

Oxidised variants of monoclonal antibodies (MAbs) can be analysed by hydrophobic interaction chromatography (HIC). In this application note, the separation of MAb samples and their oxidised species using YMC’s HIC column, BioPro HIC BF, is described. BioPro HIC BF is a more hydrophobic HIC phases dedicated for proteins of low hydrophobicity and oxidised MAb variants.

MAb oxidation with t-BHP treatment

tert-Butyl hydroperoxide (t-BHP) was used as a chemical oxidant to promote oxidation of methionine residues of NIST MAb. Subsequently, papain was used for the preparation of two Fab fragments and one Fc fragment.

Analysis of oxidised MAbs

When analysing the oxidised NIST MAb using the ammonium sulphate (a), four peaks appeared at earlier elution times compared to the peak of the non-oxidised MAb, presumably due to the conformational changes via the oxidisation of the methionine residues.

Applying sodium chloride (b), eight peaks appeared at earlier elution times compared to the peak of the non-oxidised MAb. The better resolution was further achieved within shorter analysis time compared to the usage of ammonium sulphate (a).

Analysis of papain-digested oxidised MAb

The papain digests of NIST MAb samples with and without oxidisation were analysed. The Fab and Fc fragments were characterised from the chromatogram of the papain digested MAb.

In the chromatogram of the papain digested oxidised MAb, multiple peaks appeared at earlier elution times compared to the peaks assigned to the Fab and Fc fragments. These peaks correspond to oxidised fragments according to a previous report [1], describing that oxidised fragments elute earlier than the non-oxidised ones.

[1] Journal of Chromatography A, 2008, 1214, 81-89


Digital Edition

Chromatography Today - Buyers' Guide 2022

October 2023

In This Edition Modern & Practical Applications - Accelerating ADC Development with Mass Spectrometry - Implementing High-Resolution Ion Mobility into Peptide Mapping Workflows Chromatogr...

View all digital editions

Events

AOCS Annual Meeting & Expo

Apr 28 2024 Montreal, Quebec, Canada

SETAC Europe

May 05 2024 Seville, Spain

ChemUK 2024

May 15 2024 Birmingham, UK

MSB 2024

May 19 2024 Brno, Czech Republic

Water Expo Nigeria 2024

May 21 2024 Lagos, Nigeria

View all events