The chirality of a drug can potentially have a large impact on its biological activity, metabolism and toxicity. Obtaining optically pure compounds has become increasingly important in the discovery of therapeutic compounds since one enantiomer can have positive therapeutic properties while the other can display non-therapeutic or negative biological activity. Due to the often difficult and limiting nature of achiral synthesis, chiral chromatography is typically applied to access stereochemically pure compounds. The application of Supercritical Fluid Chromatography (SFC) to chiral separations has proven very effective in recent years due to its many advantages over chiral HPLC, such as shorter retention times, higher efficiencies per unit time, and the reduction of organic solvent waste. At Abbvie, the Analytical and Purification Sciences (APS) group has provided a chiral preparative SFC service since 2000 [1, 2], which has steadily grown to impact over 30 projects and over 200 samples per year. Recently, the APS group at AbbVie has developed strategies to meet the growing need and variability of chiral separations within Drug Discovery, such as developing a streamlined scale-up approach, applying structure similarity software to minimise method development, and capitalising on the powerful stacked injection capability of SFC to provide a 3-5 day turnaround for chiral separations [3].

Impact and Discussion
At AbbVie, SFC is the technique of choice for all chiral separations supporting Discovery. SFC provides higher diffusivity of analytes in the mobile phase and can handle faster flow rates, resulting in higher efficiency over HPLC.  In addition, the use of CO2 and alcohol modifier, such as MeOH, provide SFC with a green advantage, since shorter run times with only a small percentage of organic solvent in the mobile phase results in less organic solvent consumed and waste generated.  SFC also permits rapid drydown of wet fractions without the use of aqueous solvents in the mobile phase.  For chiral separations, where samples are generally less complex than typical medicinal chemistry reverse-phase purifications and where gram-scale separations are often requested, these advantages make SFC the preferred choice in the purification laboratory. At AbbVie, our chiral separation capabilities have evolved to handle chiral separations in high throughput by implementing a variety of SFC instrumentation to meet the demand for chiral separations at the milligram to 10s of grams scale, developing a custom column-screening analytical SFC to screen 20 columns in a single login, and by employing spectral data handling software from ACD/Labs to do structure similarity searches to reduce chiral stationary phase (CSP) screening for similar analogues.

Analytical method development is performed within 24 hours of receiving the sample with reports emailed to the client and also linked directly to their ELN. Three analytical Aurora/Agilent SFC instruments are employed to meet the screening and method development needs of the chiral separation service. To effectively screen up to 20 unique CSPs a login wizard was developed using Agilent macros to control column-switching valves and build a sequence table (Figure 1).

Figure 1. Custom Aurora/Agilent Column Screening Macro. The macro is able to screen up to 20 unique CSPs against 10 columns. The runs are typically 8-10 minutes in length and all data is emailed and archived to electronic laboratory notebooks.
To View this figure and other graphics you need to download the article PDF by unlocking it.

Free to read

This article has been unlocked and is ready to read.

Download


Digital Edition

Lab Asia 31.2 April 2024

April 2024

In This Edition Chromatography Articles - Approaches to troubleshooting an SPE method for the analysis of oligonucleotides (pt i) - High-precision liquid flow processes demand full fluidic c...

View all digital editions

Events

AOCS Annual Meeting & Expo

Apr 28 2024 Montreal, Quebec, Canada

SETAC Europe

May 05 2024 Seville, Spain

InformEx Zone at CPhl North America

May 07 2024 Pennsylvania, PA, USA

ISHM 2024

May 14 2024 Oklahoma City, OK, USA

ChemUK 2024

May 15 2024 Birmingham, UK

View all events