• Determination of optimum method parameters for the analysis of oligonucleotides via HILIC
    Figure 1: Analysis of DNA oligonucleotides using the bioinert coated YMC-Accura Triart Diol-HILIC (blue) and the corresponding stainless-steel column (grey).
  • Figure 2: Influence of mobile phase pH on the separation of an all PO RNA mixture using an acidic (red), a neutral (blue) and basic pH (green).

Chromatography

Determination of optimum method parameters for the analysis of oligonucleotides via HILIC

Sep 29 2023

Due to the highly polar nature of oligonucleotides, hydrophilic interaction liquid chromatography (HILIC) is ideally suited for their characterisation. In this technical note, the separation of mixtures of short DNA (dT10, dT15 and dT20) and short RNA oligonucleotides (14, 17, 20 and 21mer) was used to examine important parameters for method optimisation in HILIC mode.

Influence of hardware

Oligonucleotides can be irreversibly adsorbed onto the metallic surface of the column hardware, due to their electron-rich backbone. The use of a bioinert coated YMC-Accura Triart Diol-HILIC column shows better peak shapes and higher sensitivity due to virtually no adsorption compared to the corresponding stainless-steel column (see Figure 1).

Influence of sample solvent and mobile phase pH

The influence of the sample solvent on peak shape especially for early eluting peaks is significant. As water is the strong eluent in HILIC mode, the ratio of organic solvent must be the same or higher than the initial gradient composition. The analysis of the DNA oligonucleotides diluted with a lower organic ratio (40%) shows that a higher injection peak as well as early eluting additional peaks occurs.

The mobile phase pH has a significant effect on retention and peak shape. Figure 2 shows that oligonucleotides are more strongly retained when the pH becomes lower. But with an acidic pH massive adsorption of the oligonucleotides occurs. Using the basic mobile phase, the retention is the lowest, but the resolution is visibly reduced, so that the neutral mobile phase pH shows the best results for this mixture.

Summary

These findings result in the following recommendations:

  • Bioinert hardware provides optimum peak shapes and high recovery
  • High organic ratio of the sample solvent is important
  • Neutral to basic mobile phase pH is recommended

You will find further information about oligonucleotide analyses in YMC’s Oligonucleotide columns brochure.


Digital Edition

Lab Asia 31.2 April 2024

April 2024

In This Edition Chromatography Articles - Approaches to troubleshooting an SPE method for the analysis of oligonucleotides (pt i) - High-precision liquid flow processes demand full fluidic c...

View all digital editions

Events

SETAC Europe

May 05 2024 Seville, Spain

InformEx Zone at CPhl North America

May 07 2024 Pennsylvania, PA, USA

ISHM 2024

May 14 2024 Oklahoma City, OK, USA

ChemUK 2024

May 15 2024 Birmingham, UK

Water Expo Nigeria 2024

May 21 2024 Lagos, Nigeria

View all events