Signal-to-Noise Spec Replaced with Instrument Detection Limit for Triple Quadrupole GC/MS

Mass spectrometry & spectroscopy

Signal-to-Noise Spec Replaced with Instrument Detection Limit for Triple Quadrupole GC/MS

20 Feb, 2012

Published over 14 years ago. See the latest and most current information on Mass spectrometry & spectroscopy.

Agilent Technologies, Inc announced that it is replacing the signal-to-noise specification with a new instrument detection limit specification for its 7000B triple quadrupole gas chromatography/mass spectrometry system. “Twenty or even 10 years ago, signal-to-noise was a reasonable indicator of GC/MS performance,” said Terry Sheehan, PhD, Agilent GC/MS Marketing Manager, “But GC/MS/MS baseline noise for simple standards is often too low and too inconsistent to be meaningful. Selection of different baseline segments can change the signal-to-noise ratio by five or even tenfold. Our new specification, the instrument detection limit, is based on the system’s precision, and precision directly correlates with ion count – the real measure of MS sensitivity.” Agilent considers IDL a better measure of GC/MS/MS performance because it follows the guidelines of the International Union of Pure and Applied Chemistry, the United States Environmental Protection Agency and many other organisations. IDL calculates typical

performance from a series of automated injections for an accurate view of system performance. The well-established Student’s t-test is applied to eight consecutive injections of 100 fg OFN using a 99% confidence interval. Additionally, the IDL specification confirms the performance of every component in the GC/MS/MS system, from the autosampler through the detector. The IDL specification of the Agilent 7000B GC/MS/MS system in EI mode is 12 fg OFN.“Dividing the MS signal of OFN standards by a noise value approaching zero gives a signal-to-noise ratio in the thousands, but that has no correlation to the system’s true performance,” Sheehan said. “It is time for ultra-low noise MS systems to use statistical detection criteria that are similar to the statistics that MS operators use for their methods.”

Latest News

Lab Asia 33.4 - August 2026

Explore our Digital Edition

Discover the latest news and research

Digital edition

Explore Our Other Sites

Envirotech Online
New oxygen sensor range targets OEM gas detection applications
Explore more Arrow
Pollution Solutions Online
Skid-mounted systems target oxygen and moisture removal for renewable natural gas
Explore more Arrow
Petro Online
Automated freezing-point testing arrives for jet fuel: what ASTM D8615 changes for lab workflows
Explore more Arrow
Chromatography Today
UHPLC–MS/MS methodology detects multiple algal toxins in fish plasma
Explore more Arrow