Laboratory Products

Specifications of Ovens Verified by Temperature Distribution Measurement - Tim Tröndle

Nov 29 2010

Author: Tim Tröndle on behalf of Testo Ltd

Free to read

This article has been unlocked and is ready to read.

Download

No direct conclusions about the reliability of the control behaviour while heating up can be drawn from the technical data on temperature stability and homogeneity as provided by the manufacturers of ovens and incubators. For this reason, the accredited service provider Testo Industrial Services was contracted to test the Memmert UFP 500 oven. The oven yielded top results for temperatures of 50°C, 150°C and 250°C and fan speeds from 0 to 100%.

An increasing number of manufacturers of ovens and incubators now proudly claim in their data sheets that their appliances are tested in accordance with DIN 12880, which has been tightened in 2007. In reality, yet all measurements were done once and do not neccesarily allow conclusions to be made about the control behaviour of an appliance over longer periods of time.

In order to reproduce processes at any given time in daily routines, however, reliability in temperatrure stability and uniformity are imperative. constant temperatures and controlling accuracy are imperative. Shortly after the introduction of the new standard, the accredited calibration lab Testo Industrial Services therefore conducted
measurements on the Memmert UFP 500 oven in accordance with the updated 12880:2007-05. These measurements clearly showed that in spite of comparable
statements of other manufacturers with respect to temperature distribution and stability, there were significant temperature deviations while heating up and in a stabilised state over longer periods of time. This indicates that several appliances on the market are in dire need of having their heating and controlling technology optimised.

Free to read

This article has been unlocked and is ready to read.

Download


Digital Edition

Lab Asia 31.2 April 2024

April 2024

In This Edition Chromatography Articles - Approaches to troubleshooting an SPE method for the analysis of oligonucleotides (pt i) - High-precision liquid flow processes demand full fluidic c...

View all digital editions

Events

ISHM 2024

May 14 2024 Oklahoma City, OK, USA

ChemUK 2024

May 15 2024 Birmingham, UK

Water Expo Nigeria 2024

May 21 2024 Lagos, Nigeria

Discovery Europe 2024

May 22 2024 Basel, Switzerland

View all events