Metrohm presents a system for fully automated inline eluent preparation. Based on Metrohm’s intelligent ion chromatography systems,
the new technique eliminates error-prone and labour-intensive eluent preparation steps while improving the accuracy of the chromatographic analysis. Using the 800 Dosino and the 849 Level Control as the only additional devices, Metrohm’s intelligent IC systems – the 850 Professional IC and the Compact IC family – can be easily extended to perform any inline eluent preparation. Controlled by the MagIC Net™ software, the 849 Level Control monitors the eluent level while the 800 Dosino performs all dosing and liquid handling tasks.
In a test, consecutive injections of a 250-μg/L standard over approximately 20 days revealed an excellent retention-time stability. After more than 800 consecutive injections, relative standard deviations for standard anions and cations were smaller than 0.55 and 0.41% respectively. In the case of a 24-hour sequence, retention-time precision for anions and cations was better than 0.09 and 0.08%, respectively. In short, Metrohm’s inline eluent preparation system increases the retention-time reproducibility and allows the continuous determination of anions and cations without manual eluent preparation.
Digital Edition
Lab Asia 31.2 April 2024
April 2024
In This Edition Chromatography Articles - Approaches to troubleshooting an SPE method for the analysis of oligonucleotides (pt i) - High-precision liquid flow processes demand full fluidic c...
View all digital editions
Events
May 21 2024 Lagos, Nigeria
May 22 2024 Basel, Switzerland
Scientific Laboratory Show & Conference 2024
May 22 2024 Nottingham, UK
May 23 2024 Beijing, China
May 28 2024 Tel Aviv, Israel