Polynuclear Aromatic Hydrocarbons In Water By Automated Solid Phase Extraction and HPLC Analysis

Chromatography

Polynuclear Aromatic Hydrocarbons In Water By Automated Solid Phase Extraction and HPLC Analysis

04 Sep, 2013

Published over 12 years ago. See the latest and most current information on Chromatography.

FMS, Inc
2 min read
Download

Introduction

Produced as the byproducts of fuel com-bustion, polynuclear aromatic hydrocarbons (PAHs) consist of fused aromatic rings. As pollutants, they are of concern because some polynuclear compounds have been identified as carcinogenic, mutagenic, and teratogenic.  PAHs are lipophilic and therefore found in the environment, primarily in soil, sediment, and oily substances. They also appear in surface and ground water, indicating a source of pollution.

Instrumentation and Consumables

  • FMS, Inc. TurboTrace™ SPE system (solid phase extraction)
  • FMS, Inc. SuperVap™ Concentrator
  • FMS, Inc. 1 gram PAH C18 cartridge
  • Waters Alliance 2695 HPLC, UV254

Method Summary

TurboTrace SPE

  1. Condition cartridge with 15 mL dichloromethane
  2. Condition cartridge with 15 mL methanol
  3. Condition cartridge with 35 mL water
  4. Load 1 liter of water sample
  5. Dry cartridge for 20 minutes
  6. Rinse sample bottle with 35 mL dichloromethane
  7. Elute sample from cartridge with
    bottle rinse
  8. Elute sample with additional 25 mL dichloromethane

SuperVap Concentrator

  1. Pre-heat temp:  60 ºC
  2. Pre-heat time:  20 minutes
  3. Heat in Sensor mode:  60 ºC
  4. Nitrogen pressure:  9 PSI

Procedure

Three, one liter samples were acidified to pH 2 with 6N hydrochloric acid and spiked with 5 μL of Restek 31458 MA EPH aromatic hydrocarbon solution. Samples were then loaded onto the FMS TurboTrace SPE system using vacuum adjusted to -10 psi and passed across an FMS PAH C18 cartridge. After sample loading, the cartridges were automatically dried using a stream of nitrogen until no residual water remained. The cartridges were subsequently eluted using dichloro-methane from both the automatic sample bottle rinse and from the elution solvent reservoir. The extracts were concentrated to 1.0 mL and 5 mL acetonitrile was added for solvent exchange. The sample was concentrated to a final volume of 1 mL for HPLC analysis.

FOR RESULTS AND CONCLUSION - UNLOCK AND DOWNLOAD

Latest Articles

ILM 51.5 July 2026

Explore our Digital Edition

Discover the latest news and research

Digital edition

Explore Our Other Sites

Envirotech Online
New civil penalties and real-time monitoring raise the bar for water company compliance
Explore more Arrow
Pollution Solutions Online
Energy efficiency first: Why shipping must act now while low-GHG fuels scale
Explore more Arrow
Petro Online
Fifty years on, the EU reaffirms its commitment to Seveso-level industrial safety oversight
Explore more Arrow
Chromatography Today
Unlock high-resolution analysis of therapeutic oligonucleotides
Explore more Arrow