Power is not the same as Efficiency. What needs to be considered when comparing cooling heating or circulators

Mar 29 2018

Author: Michael Sauer on behalf of Peter Huber Kältemaschinenbau AG

Free to read

This article has been unlocked and is ready to read.

Download

Many research tasks in the laboratory require exact temperature control. However, choosing the right circulator is not always straightforward. Advertising claims such as ‘extremely fast’ or ‘high cooling capacity’ are not very meaningful and do not permit any objective assessment. Also difficult: the comparison of manufacturer information since the measurement methods used to determine it often differ.

It is essential for the user to be able to make meaningful comparisons. DIN 12876 defines various characteristics and measurement methods for this purpose. These should be used as an orientation for heating circulators and cryostats. Characteristics recorded in this manner facilitate a reliable comparison of the performance.

Different Concepts

There are two types of temperature control devices on the market. Along with open bath circulators, closed temperature control systems are also available, sometimes referred to as process control circulators. The latter device class originated more than 25 years ago with the ‘Unistat Tango’. The Unistat Tango concept, which was completely new back then, created a huge technological advance in temperature control. In contrast to conventional bath circulators, for the very first time the ‘Unistat Tango’ worked with a closed temperature control circuit. Unistats are circulators thermostats without an internal bath. An expansion tank replaces the conventional bath to accommodate thermal expansion and this is where the volume change takes place. The mass to be thermoregulated is reduced which in turn increases the speed of temperature change. These properties make Unistats predestined for temperature control tasks in process technology and process engineering such as in reactors, autoclaves, miniplant/pilot plants, reaction blocks and calorimeters.

Thermodynamics – how fast is a system?

When considering the question of the dynamics of a temperature control device, the heating or cooling capacity (kW) is normally used as a comparative value. The power generated in the thermostats however is not sufficient for making a meaningful assessment. The system mass which has to be thermoregulated must be considered. That is why to make a meaningful comparison the cooling power density (watts/litres) as per DIN 12876 is most suitable. Always applicable: the greater the cooling power density, the more dynamic (faster) the thermostat can react to a temperature-change requirement.

Example:
Consider two temperature control devices from different suppliers. The cooling capacity of both devices is the same; likewise the pump flowrate (l/min) and both temperature control devices are connected to identical applications (e.g. glass reactors). To be able to make a statement on the dynamics (cool-down/heat up time), we can use the following formula:

P = m * c *dT/dt
(P = Power; m = Total mass; c = Spec. heating capacity; dT = Temperature difference; dt = time)
Rearranged for the cooling time dt: dt = m * c * dT/P

The expression c * dT/P is the same for both applications. However, it’s worth taking a closer look at the mass.
Temperature-control device 1 has a mass of 5 kilograms (fill volume, not the inherent weight of the device). Temperature-control device 2 has a mass of 10 kg. The mass of the external applications amounts to 5 kg. The first case results in a total mass of 10 kg (internal filling volume plus the external application); in the second case 15 kg have to be cooled (or heated). The ratio is 2:3, or in other words: Temperature control device 1 is only needed 2/3 of the time. That means the time saving is 33%.
The example demonstrates that the cooling capacity is certainly an important factor. However, this should also be applied to the temperature control fluid volume being used. The result is the cooling capacity density and this can be used to make a meaningful comparison (see DIN). By the way: Temperature-control device 1 also saves 1/3 of the temperature-control fluid and the energy!

Pressure or Flowrate?

The flow rate of the circulation pump is another important criterion. It has a strong influence on the heat transfer between a circulator and the temperature control fluid. According to DIN, the cooling capacity should be measured at full pump output. When the pump output is reduced, the heat transfer is lower. This leads to more net cooling capacity and facilitates lower temperatures. Important for most applications is not the pressure output (bar) but the highest possible flowrate (l/min). The point is ultimately not how much power a thermostat generates; rather it is the efficiency of the heat transfer to or from the process. Output is useless if it is not efficiently transferred to the application.
Circulation pumps are ideally designed so that a turbulent flow generates a large heat transfer coefficients (alpha value) and thus efficient heat transfer to/from the internal heat exchangers (evaporator and heating). In practice, in 90% of all applications glass is used (reactors, distillation apparatus, etc.) and the permissible system pressure in these applications is maximum 0.5 bar. This is why Unistat pumps offer large circulation volumes instead of high pressure, they need significantly less (pumping) motor output to accomplish high flowrates.

Summary

When selecting a circulator there are several points that need to be considered. Ultimately, it is not only the heating and cooling capacities stated in the manufacturer’s catalogue that are relevant - rather criteria such as the cooling density, flowrate and if available case studies should be paramount since that is the only way to be able to make a realistic assessment of the heat transfer efficiency.

Free to read

This article has been unlocked and is ready to read.

Download


Digital Edition

Lab Asia 31.2 April 2024

April 2024

In This Edition Chromatography Articles - Approaches to troubleshooting an SPE method for the analysis of oligonucleotides (pt i) - High-precision liquid flow processes demand full fluidic c...

View all digital editions

Events

FORUMESURE

Apr 22 2024 Marrakech, Morroco

Korea Lab 2024

Apr 23 2024 Kintex, South Korea

Korea Chem 2024

Apr 23 2024 Seoul, South Korea

Lab Indonesia

Apr 24 2024 Jakarta, Indonesia

View all events