Protein quantitation  assay according  to Bradford with BRAND Liquid Handling Station and Roti ®-Quant

Liquid handling

Protein quantitation  assay according  to Bradford with BRAND Liquid Handling Station and Roti ®-Quant

21 Mar, 2017

Published over 9 years ago. See the latest and most current information on Liquid handling.

Introduction

The Bradford assay, developed by Marion M. Bradford (Bradford 1976), is a common spectroscopic analytical procedure used to measure the concentration of proteins in a solution.

The binding of the dye Coomassie Brilliant Blue G-250 to a protein causes a shift in the absorption maximum of the dye from 465 nm to

595 nm. The increase of absorbance at 595 nm is proportional to the amount of bound dye, and thus (over a wide range) to the concentration of protein present in the sample.

The aim of this Technical Note is to compare the manual and automated Bradford assay in terms of precision and reproducibility.

Material & Methods

According to the Instruction of Roti®-Quant in 96-well microplates, dilution series of the protein standard BSA were prepared manually with a Transferpette® S 20-200 µl as well as automatically with a single channel liquid end (10-200 µl) of the BRAND Liquid Handling Station (n=8). We modified the Roti®-Quant protocol by using phosphate buffer (50 mM, pH 6.5) instead of distilled water. While the homogenisation of the BSA solution was manually reached by vortexing, it was automatically reached by mixing.

The Roti®-Quant solution was prepared manually for both assays according to the Roti®-Quant protocol and then added to the BSA solutions with a Transferpette® S -8 30-300 µl or a multichannel liquid end (20-300 µl) of the Brand Liquid Handling Station respectively. After incubating the solutions for 5 minutes the OD595 values were measured with the multimode microplate reader Infinite® 200 pro from Tecan, plotted and compared.

Results

The results of the Bradford assay obtained with the Brand Liquid Handling Station are similar to those obtained using the classic manual protocol (Figure 1).

Conclusion

This Technical Note shows that the classic manual Bradford assay with Roti®-Quant can be performed reliably and precisely with the Brand Liquid Handling Station with similar results. The Brand Liquid Handling Station offers the possibility to automate pipetting work and thus save time during assay preparation.

Latest News

ILM Guide 2026/27

Explore our Digital Edition

Discover the latest news and research

Digital edition

Explore Our Other Sites

Envirotech Online
New IMO guidance drives demand for advanced gas detection on ships
Explore more Arrow
Pollution Solutions Online
DNV introduces new framework for measuring onboard carbon capture performance
Explore more Arrow
Petro Online
From specification to certification: Ensuring jet fuel conformity with reliable flash point testing
Explore more Arrow
Chromatography Today
Affordable liquid chromatography solvent delivery pump
Explore more Arrow